Advanced Laser based Particle Counter

AvCount AvCount Particle Counter - ASTM D7619; IP 565; DEF STAN 91-91 & 91-86

ASTM D7619, IP 565 - Test methods for sizing and counting particles in light and middle distillate fuels by automatic particle counter.

  • Fully automatic operation
  • Particle size 4μm to >70μm
  • Cumulative & distributive particle numbers
  • ISO4406: 1999 codes
  • Uses "Straight from the Bottle" samples
  • Precise and reliable results
  • Laboratory or portable field use
  • Calibrated in accordance with ISO 1117

AvCount is a new laser based particle counter for checking and maintaining the quality of aviation jet fuel throughout the distribution network.

Fuel types include aviation turbine oils, biofuels and biofuel blends, heating oil, gas oil, diesel - automotive and marine, gasoline and kerosine.

For Aviation Jet Fuel:
IP 565
Determination of the level of cleanliness of aviation turbine fuel - Portable automatic particle counter method

DEF STAN 91-91
Seta ‘AvCount’ Particle Counter IP 565 is now included in DEF STAN 91-91 Aviation Jet Fuel Specification, Issue 6 for Certifying Jet A-1. Click here for information about IP 565 in DEF STAN 91-91.

Maintaining Fuel Quality in the supply and distribution systems is a major concern. The potential for particulate (dirt) contamination is high, particularly from:

  • Refinery Processing Materials
  • Corrosion & Rust in Pipes and Tanks
  • Airborne Solids
  • Degraded & Damaged Hoses and Filters
  • Microbiological Growth

The avcount particle analyser is a rugged and portable instrument specifically designed for testing the particulate content of aviation jet fuel for particles ranging from 4μm to >70μm diameter.

The avcount uses the light obscuration principle, where the fluid under test is passed through a cell comprising a laser and a calibrated sensor. Particles in the fluid cast a shadow onto the sensor, which determines the area / diameter and number of particles in the sample.

Sampling and measurement is simple; fuel is drawn from a sample container or directly from the supply line. Avcount has a 10 bar standard pressure rating or can sample up to 315 bar line pressure using the optional pressure reducing system. An integral pump automatically draws the sample through an internal measuring chamber, removing the need for an external sampling pump.

Avcount is fully automatic once the particle count test sequence is started. The instrument flushes the cell with sample prior to commencing the measurements. A precise volume of sample is then analysed. The flushing / sample analysing sequences are automatically repeated in accordance with the selected test method.

Results are displayed on the LCD screen in real time and printed out on the integral printer as the test progresses. The results are shown for 6 particle size bands in Particles/ml and ISO Code. Results can be saved to memory for trend analysis of samples, printed out on the integral printer or downloaded to a PC or external printer.

The "Log and Show" software option allows data to be downloaded from the instrument to a PC in different formats and allows the instrument to be fuly controlled and configured from any Windows interface.

Bio-Fuel testingBio-fuel testing
Optional Software AvailableOptional Software Available
Technical Datasheets AvailableTechnical Datasheets Available

 

 

Request a quote and to find your local representative


AvCount Laser Particle Counter for DEF STAN 91-91, IP 565

Learn about the NEW AvCount2 Particle Counter

Watch the AVCount Demo Video here

Demo Video

  • More Information

  • Who should use AvCount?

    Refinery Fuel Quality/Laboratory Personnel testing fuel to DEF STAN 91-91 to release Jet A-1.

    Aviation Fuel QA personnel in laboratory and field testing Deisel Fuel environments benefit from a far superior means of ensuring fast and reliable particulate test results than is currently available.

    Fuel suppliers can provide customers with a more comprehensive and secure ‘fuel cleanliness’ report whilst significantly reducing the effort needed for testing.

    Fuel filter maintenance personnel can more easily identify and isolate any potential problem, supporting reduced maintenance costs and downtime.

  • Sample presentation for measurement

    Correct sample presentation is key to achieving consistent results. The samples must be gently rotated to ensure an even dispersal of particulate matter, but care must also be taken to avoid the creation of air bubbles that may distort the results. Seta offer two specifically designed tumbler systems for the purpose of providing a uniform representative sample, minimising the risk of creating these bubbles.

    Benchtop systemBenchtop system
    Designed for up to 500ml bottles, available with optional fittings to allow a range of bottle types.

    Floor StandingFloor Standing
    Designed for 5 litre IATA cans. IATA cans are both heavy and bulky and this unique tumber allows safe and consistent rotating of the sample.

    pdfTri-fold leaflet

  • Spares and Accessories

    Optional Accessories
    PART NUMBER DESCRIPTION
    99320-2 Ultrasonic bath KC2
    SA1002-0 Flight Case
    SA1003-0 PC Software
    SA1004-0 250ml Bottle & Caps (60)
    SA1005-0 IATA Container Sampling Kit
    SA1008-0 High Pressure Adaptor
    SA1010-0 Bottle Tumbler
    SA1014-0 500ml Bottle Carrier
    SA1016-0 Adjustable BottleCarrier
    SA1020-0 IATA 5 Litre Tumbler
    SA1022-0 AvCount Sampling Kit
    SA1030-0 AvCount Service Kit
    SA1035-0 AvCount Tool Kit
    Suggested Spares
    PART NUMBER DESCRIPTION
    SA1000-001 Sample Probe
    SA1000-002 AvCount replacement Sample Tube & Stopper Assembly
    SA1000-003 Printer Paper (10)
    SA1000-004 AV Count Exit Tube Assy
    SA1000-018 Micron Filter Tube Assy
    Calibration, Verification and Training
    PART NUMBER DESCRIPTION
    SA1001-0 Calibration Fluid 250ml
    SA1006-0 Verification Fluid 250ml
    SA1100-0 Service & Calibration

  • Specifications

    Specifications
    Particle Size Range: 4μm to >70μm
    Measuring Channels: 6 channels with simultaneous measuring capability
    Size Bands - IP 565 4μm(c), 6μm(c), 4μm(c), 21μm(c), 25μm(c),
    30μm(c)
    Size Bands - ASTM (Draft) 4μm(c), 6μm(c), 14μm(c), 21μm(c), 38μm(c),
    70μm(c)
    Results: Cumulative and distributive particle number/ml
    Volume percent distribution
    ISO Codes to ISO 4406: 1999
    Measured Sample Volume: 10ml
    Total Sample Volume Required: 80ml minimum (includes rinse cycles)
    Input (sample) Pressure: Operation up to 10 bar (optional 315 bar)
    Display and Control: LCD display, 4-key-operation
    Measuring Mode: Single or continuous measurement with a cycle period
    of 1 - 99 min
    Alarm level: User defined
    Printer: Integral thermal printer
    Computer interface: RS232C
    Voltage; 100/230VAC, 50/60Hz, max 30W
    Size (HxWxD): 37 x 33 x 21cm
    Weight: 11kg

 

Automatic operation & easy to follow menu system